Quick Spin Operation
1.
Prepare the unit for centrifugation as in the General Operation (section above Steps 1-8).
2.
Press and HOLD QUICK SPIN button (rotor should spin up to the set speed).
3.
The timer will begin counting.
4.
Once enough time has elapsed for your needs, release the QUICK SPIN button. The rotor will
spin down, stop and the unit will emit a beep when the cycle is complete.
5.
To open lid, press the OPEN button.
Other
Change Between RPM and RCF
Prior to a spin cycle, press both the SPEED UP & DOWN buttons simultaneously.
The speed setting will then convert to the other.
Quick Stop
If for any reason you need to stop the cycle quickly, press and hold the START/STOP
button. The cycle will stop more rapidly than normal. Caution The tube contents may be
disturbed during the quick stop process.
Error
If an error happens, the unit will beep and the display will indicate the error.
Press Start/Stop to clear the error from the display.
STATUS
Rotor Lock
Error Status
Motor Overload
User Stop
Balance
Temperature
Excessive Tilt
Lid Fail
Rotor Lock
ERROR
Resolution
Something is interfering with the rotor. Clear the rotor and reset.
The operator has held down the START/STOP and implemented a
quick stop.
Inspect the tubes for equal tube fill or improper placement. Once
corrected, rerun.
If the balance error continues to happen, remove the tubes and
determine if the balance error still persists with an empty rotor.
Inspect the rotor for improper installation.
The unit has exceeded the normal operating temperature.
Turn off the unit and allow to cool.
The unit has experienced a non-normal tilt event.
Make sure the unit is placed on a level surface. Once corrected,
rerun
The lid has opened during the cycle. Check for proper operation of
the lid lock mechanism. The lid should stay locked during the entire
cycle.
The unit has experienced a problem with the rotor.
Correct the rotor interference. Once corrected, rerun.
Status of Unit
Error Type
9