• When you press the memory but-
ton M again, the last individual
measurement is displayed (in this
example, measurement 03).
• When the memory button M is pressed again, you
can view your individual measurements.
• To switch the device off again, press the START/
STOP button
.
You can exit the menu at any time by pressing the
START/STOP button
• To clear the memory of the relevant user memory,
you must first select a user memory.
• Start the retrieval of the average measured values.
A flashes on the display and the average value of
all saved measured values in this user memory is
displayed.
• Press and hold the memory button M and the
START/STOP button
on the user memory you are in.
.
for 5 seconds, depending
All values of the current user memory
are deleted and the device switches
off.

7. Error messages/troubleshooting

In the event of errors, the error message
the display.
Error messages may appear if
• the systolic or diastolic pressure could not be meas-
ured (
or
appears on the display)
• the systolic or diastolic pressure was outside the meas-
urement range ("Hi" or "Lo" appears on the display)
• the cuff is fastened too tightly or too loosely
(
or
appears on the display)
• the blood pressure is higher than 300 mmHg for more
than 1.5 seconds (
• pumping up takes longer than 180 seconds (
pears on the display)
27
_ appears on
appears on the display)
ap-