2 Safety instructions

Before you install the HTP1, read through this operating manual carefully. If the
instructions contained within it are not followed, especially the safety guidelines, this
could result in danger for people, the environment, and the device and the system it
is connected to.
The HTP1 corresponds with state-of-the-art technology. This concerns the accuracy,
the operating mode and the safe operation of the device.
In order to guarantee that the device operates safely, the operator must act compe-
tently and be conscious of safety issues.
KELLER provides support for the use of its products either personally or via relevant
literature. The customer assumes responsibility for any testing to determine applica-
tion suitablility. KELLER is not reponsible for damages due to misuse or application
incompatability.
Qualified personnel
● The personnel who are charged for the installation and operation of the HTP1 must
hold a relevant qualification. This can be based on training or relevant experience.
The personnel must be aware of this operating manual and have access to it at all
times.
General safety instructions
● In all work, the existing national regulations for accident prevention and safety in the
workplace must be complied with. Any internal regulations of the operator must also
be complied with, even if these are not mentioned in this manual.
● Never use the hand pump together with an external pressure source. Do not attach
an external pressure generator to the hand pump.
● Do not use brake fluid or other aggressive media.
● Do not remove any attached components (test item, pressure hose, reference gauge)
when the hand pump is under pressure:
– Open the pressure relief valve before removing any of the components.
– Do not use Teflon tape to seal the pressure connections. Surplus Teflon tape can
enter the hand pump and damage it.
– Only use adapters and seals that are available as accessories.
– Non-pressurised storage: Only store the hand pump with the pressure relief valve
open. This ensures that no pressure can be built up by unintentional pumping
movements.
– Avoid external force of all kinds towards the hand pump and its operating
elements.
– Do not use the hand pump if it is damaged or defective.