Putting into Operation

3.1
Connecting the Probe to the R&S RTx Oscillo-
scope
The R&S RT-ZC20B current probe has been designed for use with R&S RTx
oscilloscopes.
The output of the probe is terminated internally. Use an R&S RTx oscillo-
scope with high impedance inputs. With an impedance of 50 Ω, accurate
measurement is not possible.
Risk of circuit damage
Check that the conductor being measured is not clamped when connecting
the R&S RT-ZC20B to the R&S RTx oscilloscope. When probe power is
turned on, a demagnetizing waveform is initially applied to the core. This is
intentional in the design, and not a fault.
► Connect the probe box (1) to the Rohde & Schwarz probe interface of the
base unit (2). The probe will snap in when connected properly to the port.