Description; Design; Connections - Suction-/Pressure Side; Connecting To The Electricity Supply - Carl Roth CR-KP292 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Description

3
Description
3.1

Design

3.1.1

Connections - Suction-/Pressure side

Vacuum connection
Suction side
(1)
(2)
Exhaust connection
Pressure side
3.1.2

Connecting to the electricity supply

The Piston Pump is supplied with complete electrical wiring. It is connected via a mains con-
nection cable and a power plug. Mains connection cable and plug must comply with the re-
quirements of the line disconnection devices (current, output).
Motor is provided with a thermal overload protection ex works, protecting the vacuum pump
from damage or destruction, respectively.
Should the user change the electrical connection, for example for fitting into a sys-
tem, then this may only be performed by a electrical specialist under observance of
the accident prevention regulations.
Please keep in mind that the barometric pressure changes from day to day.
This also changes the displayed pressure values because of latent air pressure
fluctuations.
8
Suction side - Connection:
Hose nozzle DN 6 (1)
Pressure side - Connection:
Hose nozzle DN 6 (2)
WARNING !
WARNUNG !

Publicité

Table des Matières
loading

Table des Matières