HoMedics Deluxe BPA-3000-EU Manuel D'instructions page 16

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
IB-BPA3000-0912-01:Layout 1 09/01/2013 17:43 Page 16
Table 3 Guidance and manufacture's declaration – electromagnetic immunity – for ME
EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy.
a
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the ELE007839V1 is used
exceeds the applicable RF compliance level above, the ELE007839V1 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the ELE007839V1.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
16

Publicité

Table des Matières
loading

Table des Matières