6.2 Timer setting
By default, this feature will be OFF. Therefore, the equipment will keep working until it is manually switched OFF.
▪ Press the SET button. The controller will run into temperature setting state. Press the SET button again, the controller will
run into time setting state.
▪ In the time setting state, use the "▼" and "▲" buttons to select the desired time value (1-9999 min). Use the "◄"
button to go from one digit to the next. Press the SET button again, it will quit the setting state and the settings
will be saved automatically.
▪ When the temperature reaches the value set at 6.1, the timer will start. When the run time is over, the SV window
will display "End", the buzzer will make a sound for 30 seconds and the heating will be disconnected.
▪ In order to stop the buzzer sound, you can press any button from the controller.
▪ In order to switch OFF this feature, you can set the desired time value to 0000 min.
6.3 Temperature alarm
▪ The alarm is set by default at 5ºC.
▪ This means that when the working temperature is 5ºC higher than the desired temperature, the alarm will activate.
▪ When the temperature alarm activates, the buzzer sounds and "ALM" lights. To stop the sound, you can
press any button.
▪ If a 5ºC over-temperature is reached by changing manually the temperature value (6.1), "ALM" lights but
the buzzer does not sound.
6.4 Other concerns
▪ In the setting state, the controller will return to run status if no key is pressed in one minute.
▪ If the display window shows "----", it indicates the fault of temperature.