3 In the setting interface, user can set temperature, preheat function and the timer of A, B, C block. Timer of one block can be
changed even when other blocks are working.
Note: If the preheat function is "ON", device will automatically heat up to the target temperature after power is switched on.
If the preheat function is "OFF", device will only heat up to the target temperature after pressing "Enter".
Operation
1 After the device is powered on, if "Preheat" is in "ON" status, the incubator will automatically heat to the target temperature. In
case "Preheat" is off, press "Enter" to start heating. Once it reaches the target temperature, "Ready" will display on the top left
corner of screen with a beep.
2 Press the independent buttons "A Timer", "B Timer" and "C Timer" to start time counting for each block.
Example: Press "A Timer" to start counting for block A. "End" will display with 3 times long beep when time is up. During time
counting, press "A Timer" once to pause counting, press again to continue counting. Long press "A Timer" 5 seconds to exit
counting.
Operations on "B Timer" and "C Timer" follows the same method.
3 Before the target temperature has been reached, pressing "A Timer"/ "B Timer"/ "C Timer" will display "Err" with 3 times short
beep alarm.
4 In case of shut-down or unexpected power-cut, the device will release a short beep alarm.
Temperature deviation setting
To be used in case temperature shown on the device is not in line with actual temperature measured with a reference thermometer.
1 Press the "Set" and "▾" simultaneously to enter the temperature deviation setting.
2 It will display the current deviated value of the temperature on the left of the screen, and display the last-set deviated tempera-
ture on the right of the screen.
3 Press "▴" and "▾" to set the value of temperature to be deviated and press "Enter" to save and go back to the main interface. The
range of temperature deviation allowed is +/- 3°C.
4 Long press "Set" 5 seconds to eliminate the deviated value.
Note: If you don't want to set temperature deviation after entering the temperature deviation setting interface, press "Set" and "▾"
simultaneously to exit.
Troubleshooting
Error
Cause
No display
No power connec-
tion; Power switch
failure; Controller
failure
Strong Deviation
Sensor broken
display temp &
actual temp
Display "E001"
Over-heating tem-
with alarm
perature alarm
Display "E005"
Sensor open circuit Contact vendor
with alarm
Display "E006"
Sensor short circuit Contact vendor
with alarm
Display "E008"
Heater failure
with alarm
Press invalid
Keyboard failure
Technical data
Solution
Model
Plug in cable on
Power supply
both sides; Check
Power capacity
the power switch;
Average power consumption
Contact vendor
Contact vendor
Temperature range
Timing range
Temperature accuracy
Turn off the power
Display accuracy
and contact vendor
Heating time
(from 25°C to 70°C)
Ambient temperature
Dimensions Device (WxDxH)
Contact vendor
Dimensions Block (WxDxH)
Weight
Contact vendor
DC12V 5A
45 W
10 W
RT+5°C ~ 70°C
1sec ~ 99min59 sec
≤± 2,0°C
0,1°C
≤15min
5°C ~ 35°C
188mm x 227mm x 55mm
160mm x 105mm x 18mm
1,1 kg
EN