English Instructions For Use; Cleaning And Disinfecting - Philips M1900B Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5

English Instructions for Use

These cables connect a validated SpO
Philips patient monitors if the cable is listed as an accessory in
the monitor's Instructions for Use.
Before use, verify that combination of instrument/adapter
cable/extension/sensor is specified in the HP/Agilent/Philips
Instructions for Use. Unspecified combinations may result in
patient injury.
Key to illustration on page 1
1 M1940A (2m adapter)
2 M1900B (3m adapter)
3 M1941A (2m extension)
M1943A (1.1m adapter)
4
M1943AL (3m adapter)
M1943NL (3m adapter)
5
(with lavender-blue plug jacket)
TM
OXIMAX
is a registered trademark of Nellcor Puritan
Bennett Inc.

Cleaning and Disinfecting

Use only the approved agents listed. Others may damage the
cable or its connecting wires, shorten the product lifetime, or
cause safety hazards.
sensor to HP/Agilent/
2
12-pin cable to 8-pin
12-pin to 9-pin D-sub
8-pin to 8-pin
8-pin to 9-pin D-sub
8-pin to 9-pin D-sub
8-pin to 9-pin D-sub
(OXIMAX)
Never immerse or soak the cables or connectors in liquid. Do
not allow the cleaning agent or disinfectant to remain on the
cable; wipe it off immediately using a cloth dampened with
water.
If you see signs of physical or electrical deterioration or
damage, replace the cable. Do not use it for further patient
monitoring. Disinfect, decontaminate, and dispose of it
according to local laws for hospital waste disposal.
Cleaning
Use a lint-free cloth moistened with warm water (40°C/104°F
maximum) and soap, or a diluted non-caustic detergent, or a
1% salt solution.
Disinfecting
Use a lint-free cloth moistened with one of these low level
disinfectants:.
Alcohol based
Ethanol 70%, Isopropanol 70%, Microzid Liquid
Aldehyde based
®
Cidex
Clean the cable before disinfecting, and disinfect only when
necessary, as determined by your hospital's policy, to avoid
long-term damage to the cable.
Do not autoclave the adapter cable or use bleaches containing
sodium hypochlorite such as Clorox.
2
Opa Solution
®

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

M1940aM1941aM1943aM1943alM1943nl

Table des Matières