Unsatisfactory/No Desifection - Ecolab Eco-Satellite SE11 Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2.6.3 Unsatisfactory / no disinfection
Is the change over
Is the disinfection
nozzle connected?
Is the filter in hose from the
container with disinfection
Is there disinfection in the
10
No
handle in foam
position?
Yes
No
Yes
Is the air supply
No
disconnected?
Yes
Yes
clogged?
No
No
container?
Yes
Call a service
technician
Set the change
over handle to
foam position
Connect a 40/30
disinfection nozzle
Close the closing
valve for air
Clean or replace
the filter
Add disinfection to
the container