Ensure that the LED display shows 2000
seconds, then insert the Variable Timer Key
contained in the package of the catheter be-
3
ing used in the procedure. Confirm that the
LED display changes within 2 seconds after
insertion and displays the correct time value
corresponding to the key inserted.
CAUTION: FAILURE TO USE THE PROPER VARIABLE TIMER KEY CAN RESULT IN INADEQUATE CURING OF THE IL-
LUMINOSS MONOMER. ENSURE YOU ARE USING THE CORRECT TIMER KEY CORRESPONDING TO THE IMPLANT
SELECTED FOR SURGERY.
NOTE: There is a protective shroud protruding from the faceplate which prevents improper insertion of the
Variable Timer Keys. See image below for help with insertion.
Ensure that the STANDBY TOGGLE SWITCH
is in the "Foot Pedal Enabled" position,
then operate the Light Source Console by
depressing the Foot Pedal. The Light Source
4
Console will operate for the prescribed
number of seconds. With the STANDBY
TOGGLE SWITCH in the "System Standby
Mode, " the Foot Pedal is inoperable and the
user will hear an audible beep when the
foot pedal is depressed.
CAUTION: DO NOT REMOVE TIMER KEY DURING COUNTDOWN!
CAUTION: If STANDBY TOGGLE SWITCH is changed to "System Standby Mode" during operation, wait for time
cycle to expire, remove key and reinsert, then reinitiate full cure cycle.
CAUTION: This is an arc, not a filament bulb. Avoid turning the console off and on. It is recommended that
the console is turned on at the beginning of the procedure and remains in operation until the end of the
procedure.
CAUTION: The bulb must be cool before it can be reignited. Turn the unit off and allow 5 – 10 minutes for
it to cool down. If the bulb fails to reignite, refer to the "Troubleshooting" section of this manual. Bulb life is
reduced each time the unit is switched on and off. Avoid repeated cycles that shorten bulb life by leaving the
unit on throughout the procedure.
900093.S
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE - ILLUMINOSS MEDICAL INC
PAGE 9 OF 96
2000
.