The anodized aluminum housing contains an electronic power supply, circuit protection, bulb/reflector
assembly, internal light filter for extended light guide life, thermostatically controlled cooling fan, light guide
mount, bulb and unit status indicator lights, and a shutter system. A thermal shutdown sensor is provided
for internal temperature control of the unit. A cover closed switch and light guide sensing switch add to
the safety of the unit. Fan filters should be changed or cleaned frequently to prevent blockage and reduced
ventilation airflow. The electric shutter is supplied with foot pedal enabled and system standby operating
modes. The light guide is separate and plugs into the light guide holder.
WARNING:
WARNING:
The power supply operates on line voltage between 100-240 VAC and frequencies between 50-60 Hz. If the
bulb extinguishes due to a momentary power failure, the unit must be turned off, allowed to cool, and then
turned on again in order to reignite the bulb. A cooling fan is provided to keep the bulb housing and internal
components of the power supply at the optimum operating temperature. The fan must not be covered or
otherwise blocked. The light source is a 75-Watt short arc mercury vapor bulb mounted in a reflector and
pre-focused to provide optimum light output. The unit is rated for continuous operation.
The system is designed to operate in conjunction with the IlluminOss Variable Timer Keys which will be
supplied with each IlluminOss Photodynamic Bone Stabilization System. The keys contain specific resistor
circuitry which, when connected to the light box, will program a specific timer countdown cycle which is
optimized for the specific IlluminOss implant.
Testing confirms a minimum acceptable intensity provided by the lamp bulb. A "Change Bulb" Indicator light
is provided to alert the operator to check bulb operation or to contact customer service if maintenance is
required.
900093.S
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE - ILLUMINOSS MEDICAL INC.
Foot Pedal
Engage the light guide in the mount before the light is turned on, and
remove the light guide from the bezel only after the light is turned off to
avoid the accidental possibility of exposure to the light.
UV emitted from this product.
Front Panel Light Display
SYSTEM READY
LIGHT GUIDE SEATED
CHANGE BULB