7. Operation

The Photodynamic Curing System will arrive almost fully assembled. The system should be positioned in a
dry location that will not obstruct the rear of the unit. Adequate ventilation is essential for proper operation.
CAUTION: TO ENSURE PROPER OPERATION OF THE SHUTTER, BE SURE TO COMPLETELY INSERT
THE LIGHT GUIDE INTO THE CONSOLE. ENSURE THAT THE LIGHT GUIDE REMAINS IN PLACE DUR-
ING USE.
To power the system, turn the Power Switch to the ON position; the fan, timer, and 75W lamp should begin to
function. The operator should confirm proper start up before commencing with surgery by checking for an
audible fan system and visually confirming a green light shining out the back of the box. Before operating the
unit, allow the 75W lamp to complete the warm up cycle, approximately five minutes.
CAUTION: NEVER LOOK DIRECTLY AT LIGHT EXITING THE LIGHT GUIDE OR THE LIGHT PIPE.
CAUTION: IF POWER TO THE CONSOLE IS DISTRUPTED DURING A CURING CYCLE, TO ENSURE PROPER
CURING HAS BEEN ACHIEVED, IT IS RECOMMENDED THAT A FULL REPEAT CYCLE IS PERFORMED FOL-
LOWING THE RESUMPTION OF POWER AND A FULL SYSTEM RESTART. THE CYCLE TIME IS AUTOMATI-
CALLY SET BY THE TIMER KEY INCLUDED IN THE CATHETER PACKAGE.
CAUTION: WHEN THE "CHANGE BULB" LIGHT ILLUMINATES, THE UNIT WILL NOT OPERATE UNTIL A
NEW BULB IS INSTALLED AS DESCRIBED IN THE BULB REPLACEMENT PROCEDURE.
Bulb life is reduced each time it is started. To avoid premature lamp deterioration, leave the unit on through-
out breaks and short shutdowns. These light sources are designed for continuous operation. If the power is
lost momentarily, shut the unit off to allow the unit to cool down for 5 -10 minutes, then restart the console.
Turn the power switch to "ON. " The LED
1
display should run through a power up
sequence and end on 2000 seconds.
Allow the bulb to warm up (3-5 minutes).
The bulb is warmed up when the GREEN
"System-Ready" indicating light is illumi-
2
nated. The shutter is inoperable until the
warm up period is complete and the Light
Guide is installed.
900093.S
PHOTODYNAMIC LIGHT CURING SYSTEM INSTRUCTIONS FOR USE - ILLUMINOSS MEDICAL INC.