Warranty Information; Maintenance And Service; Cleaning; Fuses - Grant GLS Aqua 18 Plus Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

8.0

Warranty information

When used in laboratory conditions and according to these working instructions, this
product is guaranteed for THREE YEARS against faulty materials or workmanship.
Extended warranty for years four and five can be purchased by contacting our sales
department at labsales@grant.co.uk.
9.0

Maintenance and service

No routine maintenance is required except for cleaning. There are no user serviceable
parts inside the unit.
9.1

Cleaning

Clean the outside of the equipment with a damp cloth, using water only. Do not use
chemical cleaning agents. Before using any other cleaning or decontamination method,
check with Grant Instruments or your local representative to make sure that the proposed
method will not damage the equipment.
Scale on immersed parts can be removed using chemical de-scaling products designed
for use on kitchen equipment that have metal parts. De-scaling products may be toxic
and manufacturer‟s instructions should always be followed.
9.2

Fuses

Fuses are Littelfuse 3AB 314 series, fast-acting, high breaking current (max breaking
current at least 750 A); dimensions are 1.25 inch long, 0.25 inch diameter. Replace fuses
only by the same type and rating (250V).
Models
GLS Aqua 12 Plus
GLS Aqua 18 Plus

9.2.1 Replacing fuses on GLS Aqua Plus

Disconnect the unit from the power supply
1. Remove the mains lead from the socket at the back of the bath.
2. Press down the fuse drawer catch and pull out the fuse drawer. Check the fuse(s)
and replace if necessary using the specified fuse type and rating.
3. Push back the drawer and replace the mains lead.
29339 Ver. 3
Page 15
Fuse Rating (A)
120V
230V
10
15
www.grant.co.uk
10
15
GLS Aqua Plus