Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR
The I-TECH UE is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such an
environment
Immunity test
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR
EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
The I-TECH UE device is intended for use in. the electromagnetic environment
specified below. The customer or the user of the I-TECH UE should assure that it is
used in such an environment.
Immunity test
Portable and mobile RF communications equipment should be used no closer to any
part of the I-TECH UE device, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
ove P è la potenza massima nominale d'uscita del trasmettitore in Watt (W) secondo il
costruttore del trasmettitore e d è la distanza di separazione raccomandata in metri (m).
Le intensità di campo dei trasmettitori a RF fissi, come determinato da un'indagine
elettromagnetica
del sito potrebbe essere minore del livello di conformità in ciascun intervallo
a
di frequenza
b
Si può verificare interferenza in prossimità di apparecchi contrassegnati dal seguente
simbolo:
Note:
IACER Srl
ALL EQUIPMENT AND SYSTEMS
Test level
IEC 60601
30A/m
Test level
Compliance
IEC 60601
3V
3V
eff
eff
from
150kHz
from
to 80MHz
to 80MHz
10V/m
10V/m
from 80MHz to
[E
] V/m
1
2,7GHz
Compliance level
30A/m
level
environment – guidance
Recommended separation distance
from 150kHz to 80MHz
150kHz
from 80MHz to 800MHz
from 800MHz to 2,7GHz
58
Electromagnetic
environment -
guide
Power
frequency
magnetic
fields
should be at levels
characteristic of a
typical location in a
typical commercial
or
hospital
environment
Electromagnetic
d = 1.2√P
d = 1.2√P
d = 2.3√P
MNPG194-01