3. Operation
3.
When tooth types are selected, the exposure angles are automatically set according to
the tooth type. To check the default exposure angles, refer to the following table.
Refer to the following figure to see the angle for molar/premolar.
18
Tooth Type
Incisor
Canine
Molar/Premolar
Bitewing
Since the angles of inclination for the maxilla incisor and canine are
same, the exposure time of the canine is applied for both the maxilla
incisor and canine.
Angle of Inclination
Maxilla: +40° ~ +50°
Mandible: -22° ~ -28°
Maxilla: +40° ~ +50°
Mandible: -17° ~ -23°
Maxilla: +25° ~ +35°
Mandible: -2° ~ -8°
+3° ~ +12°