KaWe MedCenter 5000 Mode D'emploi page 21

Masquer les pouces Voir aussi pour MedCenter 5000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24
Guidelines and manufacturer declaration – electromagnetic interference immunity
The KaWe MedCenter 5000 is designed to be used in the types of electromagnetic environments listed below.
The customer or user is responsible for ensuring that this device is used in such an environment.
Interference
IEC 60601-
immunity test
test level
ELECTROSTATIC
± 6kV contact discharge
DISCHARGE (ESD) accor-
± 8 kV air discharge
ding to IEC 61000-4-2
Fast transient electrical
± 2kV for power supply
disturbances/ bursts
lines
according to IEC
± 1kV for input and
61000-4-4
output lines
± 1kV phase-to-
Surges according to IEC
phase voltage
61000-4-5
± 2kV phase-to-ground
voltage
<5% U
for 1/2 period
(>95 % dip)
Voltage dips, short
40 % U
interruptions
for 5 periods
and voltage variations
(60 % dip)
on power supply input
lines according to IEC
70% U
61000-4-11
for 25 periods
(30 % dip)
<5 % U
(>95 % dip)
Magnetic field with
a power frequency of
3 A/m
(50/60 Hz) according
to IEC 61000-4-8
Note: U
is the mains voltage prior to application of the test level.
T
Compliance
level
± 6kV contact discharge
± 8 kV air discharge
± 2kV for power supply
lines
± 1kV for input and
output lines
± 1kV phase-to-
phase voltage
± 2kV phase-to-ground
voltage
<5% U
T
T
for 1/2 period
(>95 % dip)
40 % U
T
T
for 5 periods
(60 % dip)
70% U
T
T
for 25 periods
(30 % dip)
for 5 ѕ
<5 % U
for 5 ѕ
T
T
(>95 % dip)
3 A/m
Electromagnetic environment –
guidelines
The flooring should be wood or concrete or be covered
with ceramic tile. If the floor material is a non-conduc-
tive, synthetic material, the relative humidity of the air
must be at least 30%
The quality of the mains supply voltage should be
the same as that of a typical commercial or hospital
environment.
The quality of the mains supply voltage should be
the same as that of a typical commercial or hospital
environment.
The quality of the mains supply voltage should be
the same as that of a typical commercial or hospital
environment.
If the user requires continued operation during power
interruptions, it is recommended that the device be
powered from an uninterruptable power supply or a
battery.
Should interference occur, it may be necessary to place
the KaWe MedCenter 5000 further from the source line
frequency magnetic fields or to fit it with a magnetic
shield: the line frequency magnetic field should be
measured at the intended use site to ensure that it is
not large enough to cause a disturbance.
21

Publicité

Table des Matières
loading

Table des Matières