5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with
liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
DO NOT block the venting grids of the LED light source.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light
source.
CAUTION Use only cords provided with the light source.
Plug the AC power cord into a wall outlet using the three-prong plug supplied
with the unit.
CAUTION To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power
with three-prong plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
CAUTION When used in clinical or residential areas near radio or TV units,
this equipment may be subjected to radio interference. To
avoid adverse electromagnetic effects, DO NOT operate this
equipment near RF energy equipment.
5.2 CONNECTING THE FIBER OPTIC CABLE
Turn intensity control to the minimum setting. Connect the light cable to the
instrument and then plug the light cable endfitting into the port on the front
panel.
LIT168 ZIBRA CORP
(English)
Rev. A
Page 9 of 65