Bühler technologies EGK 10 Notice De Montage page 10

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11
EGK 10
7.1 Troubleshooting
Problem /
Possible cause
Malfunction
No display
• Mains voltage in-
terrupted
• Fuse defective
Cooler doesn't
• Compressor hous-
start up
ing temperature
too high
Display flashes due to:
• Excess
• Operating point
temperat-
not yet reached
ure
• Cooling output
too low despite
the cooler run-
ning
• Flow rate / dew
point / gas tem-
perature too high
• Built-in fan
stopped
• Temperat-
• Control unit fail-
ure low
ure
Condensate in-
• Condensate trap
side the gas
full
output
• Valve inside the
automatic con-
densate drain
may be stuck
• Cooler overload
Reduced gas
• Gas circuit
flow rate
clogged
• Condensate out-
put iced over
No cooling
• Compressor
doesn't start up
Fuse is trigger-
• Increased com-
ing
pressor current
consumption due
to faulty com-
pressor start-up
10
Bühler Technologies GmbH
Error Messages in the Display
The display alternates between the temperature and error
Action
message,
• Connect to
mains; verify
Malfunction
• the plug is inser-
ted correctly
• Check fuse and
replace, if neces-
sary
• Allow to cool
down and ensure
adequate ventil-
ation
• Wait (max. 20
min)
• Ensure the vents
are not covered
(heat buildup)
• Maintain limits /
install pre-separ-
ator
• Check and re-
place if necessary
• Send in cooler
• Empty condens-
ate trap
• Flush in both dir-
ections
• Maintain limits
• Remove and
clean heat ex-
changer
• Send in cooler
• Compressor PTC
not cooled down
sufficiently. Wait
5 minutes and try
again.
• Compressor PTC
not cooled down
sufficiently. Wait
5 minutes and try
again.
Problem /
Possible cause
Error 01
• Interruption
Error 02
• Short circuit
Action
• Temperature
sensor failure:
Send in cooler
• Temperature
sensor failure:
Send in cooler
BX450005 ◦ 04/2021

Publicité

Table des Matières
loading

Table des Matières