Maintenance

Non-sterilisable
Cleaning
Use Spraynet from Bien-Air Dental fig.
9. Spray the outside of the equipment
and carefully clean the surface.
Disinfection
Disinfect using a clean cloth soaked in
a suitable product. Products containing
acetone, chlorine and bleaches are not
recommended as disinfectants. To
keep the surface of the hose in good
condition, it is advisable to periodical-
ly wipe the complete length of it with a
cloth dusted with talcum powder.
ENG
Never immerse in disinfectant solu-
tions.
O-
Not suitable for ultrasonic bath.
Only use maintenance products and
components from Bien-Air Dental. The
use of other products and components
can void the guarantee.
Transport and storage conditions
Temperature between -40°C (-40°F)
and 70°C (158°F), relative humidity
between 10% and 100%, atmospheric
pressure 50 kPa to 106 kPa (7.3 to 15.3
psi).

Other precautions for use

The device must be used by a qualified
person in accordance with the current
legal provisions concerning industrial
safety, health and accident prevention
measures, and these working instruc-
tions. In accordance with these re-
quirements, the operators:
• must only use operating devices
that are in perfect working order; in
the event of irregular functioning,
excessive vibration, abnormal heat-
ing or other signs indicating mal-
function of the device, the work
must be stopped immediately; in
this case, contact a repair centre
that is approved by Bien-Air Dental;
• must ensure that the device is used
only for the purpose for which it is
intended, must protect themselves,
their patients and third parties from
any danger, and must avoid con-
tamination through the use of the
product.
Rest the device on a suitable support
to avoid risks of infection for yourself,
the patient or third parties.
The device is not authorised for use in
an explosive atmosphere (anaesthetic
gas).
Recommendations
It is essential to use dry, purified com-
pressed air in order to ensure the long
working life of the device. Maintain the
quality of the air and the water by reg-
ular maintenance of the compressor
and the filtration systems. The use of
unfiltered hard water will lead to early
blockage of the tubes, connectors and
spray cones.

Guarantee

Terms of guarantee
Bien-Air Dental grants the operator a
guarantee covering all functional de-
fects, material or production faults.
The device is covered by this guaran-
tee for 12 months from the date of in-
voicing.
In the event of justified claim, Bien-
Air Dental or its authorised representa-
tive will fulfil the company's obliga-
tions under this guarantee by repairing
or replacing the product free of charge.
Any other claims, of whatever nature,
in particular in the form of a claim for
damages and interest, are excluded.
Bien-Air Dental shall not be held re-
sponsible for damage or injury and the
consequences thereof, resulting from:
• excessive wear and tear
• improper use
• non-observance of the instructions
for installation, operation and main-
9