TYPE
VERSION
ELECTRICAL SPECIFICATIONS
Specifications range
Common specifications
ACCURACY, THE GREATER OF GENERAL AND BASIC VALUES
GENERAL VALUES
Input type
All
BASIC VALUES
Input type
RTD
Lin.R
EMC immunity influence
Effect of supply voltage variation
Vibration
Lloyd's specification no. 1
Max. wire size
Humidity
Dimensions
Tightness (enclosure / terminal)
Weight
ELECTRICAL SPECIFICATIONS, INPUT
RTD AND LINEAR RESISTANCE INPUT
RTD type
Pt 100
Ni 100
Lin. R
Max. offset
Cable resistance per wire (max.)
Sensor current
Effect of sensor cable resistance (3-wire)
Sensor error detection
4 | 520B7737 | IC.PS.P40.A4.ML
Standard
Standard Atex
Supply voltage, DC
Internal consumption
Voltage drop
Warm-up time
Communications interface
Signal / noise ratio
Response time (programmable)
Signal dynamics, input
Signal dynamics, output
Calibration temperature
Absolute accuracy
≤ ± 0.1% of span
Basic accuracy
≤ ± 0.3 °C
≤ ± 0.2 Ω
Min. value
-200 °C
850 °C
-60 °C
250 °C
0 Ω
10000 Ω
MBT 9110
084Z7440
084Z7441
MBT 9110, 084Z7440
Standard ATEX, MBT 9110, 084Z7441
≤ ± 0.01 °C / °C
≤ ± 20 m Ω / °C
≤ ± 0.5% of span
≤ 0,005% of span / VDC
IEC 60068-2-6 Test FC
4 g / 2 – 100 Hz
1 x 1.5 mm
< 95% RH (non-cond.)
ø44 x 20.2 mm
IP68 / IP00
50 g
Max. value
Min. span
25 °C
25 °C
30 Ω
50% of selec. max. value
10 Ω
> 0.2 mA, < 0.4 mA
< 0.002 Ω / Ω
Yes
-40 °C – 85 °C
8 – 35 V
8 – 30 V
25 mW – 0.8 W
8 VDC
5 min.
Loop Link
min. 60 dB
0.33 – 60 s
19 bit
16 bit
20 – 28 °C
Temperature coefficient
≤ ± 0.01% of span / °C
Temperature coefficient
stranded wire
2
Standard
IEC 60751
DIN 43760
–