Performing the calibration check
3. Position the device as near as possible to the wall at the height of point A1.
Now adjust the device in the X axis.
!
Mark either levelling screw A or B as a reference screw. Align the unit only
with the other levelling screws.
4. Turn the device through 180° and mark point A3. The difference between points
A2 and A3 is the tolerance for the X axis.
5. To check the Y axis, repeat steps 3 and 4.
3.
A1
4.
If points A2 and A3 are more than 0.2 mm / m apart on either the X, Y or Z axis,
!
the device is in need of adjustment. Contact your authorised dealer or
else the UMAREX-LASERLINER Service Department or follow the calibration
instructions in the manual.

Calibration procedure

6. Once you have completed steps 1 to 4 and identified dots A2 (reference) and A3
(deviation), use the corresponding screws to level the laser so that dot A3 covers
dot A2. During this process, the air bubble in the level will move out of its central
position and stray to the side.
7. Remove the protective caps from the ends of the levels to
expose the calibration screws. Turn the screws to calibrate
the level (using the SW 2.5 socket wrench included in the
scope of supply). Calibration is complete when the air bubble
returns to the centre of the level.
8. Repeat for the remaining axes.
BeamControl-Master
BeamControl-Master
A2
A3
Calibration screws
A2
A2
A3
19
EN