6 Using the product
6.1.1.
Setting options
For status displays see Status displays (page 33).
1 Hz: Frequency of the movement per second (flashes per
second).
2
PULS
3
PULS
4
DELAY
the internal or external trigger signal and the flash.
5
PHASE
frequency) between the internal or external trigger signal and
the flash.
6
DIV
255.
7
OPT
trigger signal. With this option, the polarity of the trigger signal
can be defined.
•
•
36
All setting options in the display are shown in the image.
The numbering corresponds to the sequence in which they
succeed one another by pressing the
A parameter set differently than the factory setting flashes
during operation.
μs: On-time of the flash (in microseconds).
deg: On-time of the flash (in degrees).
ms: Setting of the delay time (in milliseconds) between
deg: Setting of the phase shift (in degrees, relative to
(only with external trigger signal): Pulse divider, max. value
(only with external trigger signal): Edge selection of the
0 = positive edge
1 = negative edge
[M]