Handicare MiniLift160 Classic Manuel D'utilisation page 19

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 116
Guidance and manufacturer's declaration - electromagnetic immunity
Immunity test
IEC60601-1-2 test level
Electrostatic
±2kV, ±4kV, ±8kV, ±15kV air
discharge (ESD)
IEC 61000-4-2
±8kV contact
Conducted
3V in 0.15 MHz to 80MHz
disturbances
inducted by RF
6V in ISM and amateur radio
fields
bands between 0.15 and 80
IEC 61000-4-6
MHz
80% AM at 1 kHz
Radiated RF
Home Healthcare Environment
electromagnetic
10 V/m
field
80 MHz to 2.7 GHz
IEC 61000-4-3
80% AM and 1 kHz
Proximity fields
385 MHz - 27 V/m
from RF wireless
450 MHz - 28 V/m
communications
710, 745, 780 MHz - 9 V/m
equipment
810, 870, 930 MHz - 28 V/m
IEC 61000-4-3
1720, 1845, 1970, 2450 MHz
- 28 V/m
5240, 5500, 5785 MHz - 9 V/m
Electrical fast
Signal and control lines: ±1kV
transient/burst
AC: ±2kV
IEC 61000-4-4
100 kHz repetition frequency
Power frequency
30 A/m
Magnetic field
IEC 61000-4-8
50 Hz and 60 Hz
(a) Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the MiniLift160 Classic is used exceeds the appli-
cable RF compliance level above, the MiniLift160 Classic should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary.
(b) Over the frequency range 150 kHz to 80 MHz, field strength should be less than 1V/m.
U S E R M A N U A L
Compliance level
±2kV, ±4kV, ±8kV, ±15kV air
±8kV contact
3V in 0.15 MHz to 80MHz
6V in ISM and amateur radio
bands between 0.15 and 80
MHz
80% AM at 1 kHz
Home Healthcare Environment
10 V/m
80 MHz to 2.7 GHz
80% AM and 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz - 9 V/m
810, 870, 930 MHz - 28 V/m
1720, 1845, 1970, 2450 MHz
- 28 V/m
5240, 5500, 5785 MHz - 9 V/m
Signal and control lines: ±1kV
AC: ±2kV
100 kHz repetition frequency
30 A/m
50 Hz and 60Hz
Electromagnetic environment-
guidance
Floor should be wood,
concrete or ceramic tile. If
the floors are covered with
synthetic material, the rela-
tive humidity level should be
at least 30%.
Portable and mobile RF
communications equipment
should be used no closer
to any part of the product,
including cables, than 1.0m,
if the transmitter's output
power rating exceeds 1W.
(a)
Field strength from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range.
(b)
Interference may occur in
the vicinity of equipment
marked with this symbol:
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical com-
mercial or hospital environ-
ment.
19

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

400641436400641437

Table des Matières