Télécharger Imprimer la page

Connection Of The Fore-Vacuum Pump - Agilent Technologies TwisTorr 74 FS Notice De Mode D'emploi

Masquer les pouces Voir aussi pour TwisTorr 74 FS:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 49

Connection of the Fore-Vacuum Pump

A flange KF 16 NW is available to connect the TwisTorr 74 FS pump
to the forevacuum pump. A hose or vacuum approved pipe can be
used. If a rigid pipe is used, any vibration generated by the
mechanical pump must be eliminated through the use of bellows.
NOTE
The TwisTorr 74 FS pump is characterized by its high compression ratio also for
oil vapors. When using a mechanical oil sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore vacuum pump in order to
prevent oil back streaming.
TwisTorr 74 FS User Manual / 87-901-053-01 (A.00)
Technical Information

Connection of the Fore-Vacuum Pump

17
341/352

Publicité

loading