5.2 Inserting the new saw blade

1.
Check that the female thread and saw blade support of the drive shaft are clean.
2.
Position the new saw blade. It makes no difference which side of the saw blade is
facing inwards or outwards.
Important! The notches in the saw blade must be positioned precisely flush on the
spigots of the drive shaft.
3.
Position the fixing ring. The notches in the fixing ring must always be pointing towards
the saw (see
fig. 4
4.
Screw the retaining screw in manually until hand tight.
5.
Then tighten firmly using the fork wrench, as described above. The tightening torque is
6 – 7 Nm. Caution! Do not overtighten.
22
fig. 3 Releasing the retaining screw
). These must also be positioned precisely on the spigots.
fig. 4 Inserting the saw blade
Oszillo-saw GOLD II accu autopsy