Procedure:
• Use an approx. 2mm wide flathead screwdriver.
• Find the lock screw on the laryngoscope blade. (Fig. 1)
• Loosen the screw by turning it to the left with the screwdriver.
• Check to see if the green light guide holder is now loose.
• Do not attempt to completely remove the lock screw.
EN
• As soon as the green light guide holder is loosened, slide the holder out of the slot. (Fig. 2)
• The light guide can now be detached and the curved fibre guide can be removed.
• Select a new light guide.
• Do not touch either end of the light guide.
• Insert the light guide carefully into the preformed opening in the laryngoscope blade.
Do not apply any pressure to the light guide or bend it.
• Insert the green light guide holder into the metal slot on the underside of the blade.
• Adjust the green base of the light guide holder so that it is level with the base of the
blade.
• Carefully tighten the screw until the light guide holder is tightly fastened.
• Carefully polish the input and output areas of the newly-inserted light guide.
• Test the light output intensity.
Regularly check the integrity of the light guide. In order to ensure the long-term
functionality of your product, follow the approved disinfection and sterilization
procedures.