Method Notes; Chemical Method Notes; Methods - Tintometer Lovibond MD100 Mode D'emploi

Masquer les pouces Voir aussi pour Lovibond MD100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
General notes
GB

Method notes

Prior to measurement ensure that the sample is suitable for analysis (no major interference)
and does not require any preparation i.e. pH adjustment, filtration etc.
Different Refill Packs available on request.
Reagents are designed for use in chemical analysis only and should be kept well out of
the reach of children.
Ensure proper disposal of reagent solutions.
Material Safety Data Sheets: available on request.
(Internet: www.lovibond.com)

Chemical method notes:

Method:
The organic material present in the sample is oxidised by a standard amount of a potassi-
um dichromate oxidising mixture. After oxidation is complete, the excess of this reagent is
measured photometrically.
Application:
Samples can be measured if the chloride content does not exceed 1 000 mg/l (LR/MR) or
10 000 mg/l (HR).
In exceptional cases, compounds contained in the water cannot be oxidized adequate. This
results in minimum findings, compared with the reference method.
Different methods of sampling, preparation of the sample itself and the time elapsed between
taking the sample and analysis can all affect the obtained results.
MD100_8 07/2016
19

Publicité

Table des Matières
loading

Table des Matières