Discharging Used Water; Interruption Of A Sterilization Cycle - Tecno-gaz ONYX B 7.0 Mode D'emploi

Masquer les pouces Voir aussi pour ONYX B 7.0:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 93
ENGLISH
10.8 Discharge used water
When the used water level LED (Fig. A-pos.
must be emptied.
If nothing is done, the operation of the autoclave is blocked.
Take the hose supplied (Fig.7), and insert it into the used water drain fitting located on the
front of the autoclave (Fig.A-pos.
unscrew the ring nut, turning it anticlockwise, the water will fall into the container, emptying
the tank.
IMPORTANT:
A - The hose located in the receiving container must never touch or be immersed in the
discharged water, otherwise sucking up will occur.
B - Always wait until the discharge water has been completely drained. The used water
maximum level LED turns off when some water still remains into the tank, consequently do
not considered it as a reference for this operation.
When drained, screw in the ring nut and remove the hose.

10.9 Interruption of a sterilization cycle

A sterilization cycle can be voluntarily interrupted by pressing the button
seconds.
The autoclave will send out a beep and the display will show the alarm message A001.
For reset the alarm, press at the same time, the multifunction push buttons on the
bar up to the cancelling of the same.
22
23
) lights up, the used water collection tank
12
). Put the other end of the hose in a container and
Stop
for at least 2
Reset

Publicité

Table des Matières
loading

Table des Matières