13. 4 RECOMMENDED SEPARATION DISTANCES

The device is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances
are controlled.
The device user and/or installer can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile radiofrequency communications equipment emitters and the device,
according to the maximum output power of the equipment, as recommended in the table below.
Rated max. power of the
emitter (W)
0.01
0.1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meteres (m) can be
estimated using the equation applicable to the frequency of the emitter, where P is the max. power rating of the
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These speciÀ cations may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reÁ ection from structures, objects and persons.

13. 5 CABLE LENGTHS

Cables and
Maximum
accessories
length
Handpiece
cords
Footswitch
Less than
cord
Power supply
cord
18
Separation distance in metres (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
emitter in watts (W) according the manufacturer.
RF emission, CISPR 1 - Class A/Group 1
Harmonic current emission: IEC61000-3-2
Voltage Á uctuation: IEC61000-3-3
Immunity to electrostatic discharge: IEC61000-4-2
Immunity to electrical fast transients/bursts: IEC61000-4-4
3m
Immunity to surges: IEC61000-4-5
Immunity to voltage dips, short interruptions and voltage variations: IEC61000-4-11
Immunity to conducted disturbances induced by radiofrequency À elds: IEC61000-4-6
Immunity to radiated radiofrequency electromagnetic À elds: IEC61000-4-3
80 MHz to 800 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
Complies with:
800 MHz to 2,5 GHz
d = 2.3
0.23 m
0.73 m
2.3 m
7.3 m
23 m