Adjust the irrigation Á ow rate for each programme
according to the recommendations of the tip-related
instructions (see chap. 7.4)
Press the
button to validate the new settings
(Fig. 3-4).
The storage of parameters is conÀ rmed by an audible
and visual signal (logo
Repeat this sequence to set all programmes, as
required.

7.6 TOOL KIT

Important:
The device has a T
OOL
various adjustments, such as brightness or return
to factory default settings.
To activate the T
OOL
device, wait for 4 seconds, switch the device back
on and press
and
they appear on the home screen (Fig. 6).
A quick press on button
select the brightness settings menu
the default factory settings menu
selected menu is identiÀ ed by a cross (Fig. 5-4).
a) Adjusting screen brightness
The user can adjust the brightness of the device
screen.
Select the brightness settings menu
Adjust the brightness using the buttons
(Fig. 3-1 and 2). The brightness can be adjusted
between 0% and 100%.
The brightness intensity level is indicated (Fig. 5-5).
b) Return to default factory settings
Returning to the default factory settings solely allows
irrigation Á ow rates to be restored.
Select the default factory settings menu
(Fig.5-2).
10
changes colour).
K
function used to make
IT
K
function, switch off the
IT
(Fig. 3-1 and 2) as soon as
(Fig. 3-4) allows you to
(Fig.5-1) or
(Fig.5-2). The
(Fig.5-1).
The various factory conÀ gurations will be effective
once you exit the tool kit.
The device is conÀ gured at the factory according to
the following parameters:
PROGRAMME
Very
powerful
Powerful
Average
power
Low power
c) Timer
The device has a function that indicates the duration
of use of the device. The total duration is displayed
on screen (Fig. 5-7) under the logo
The shortest duration displayed is one minute.
d) Software version
The software version is displayed at the bottom of the
screen (Fig. 5-6).
e) Storing settings
To store the modiÀ ed parameters and use the device,
hold down button
screen shown in Fig. 6 reappears.
VIII - SAFETY
and
The device is À tted with a system that can detect
malfunctions that may occur.
When an internal operating error occurs at start-up,
symbol 1 (Chap XVII - Symbols) lights up (Fig. 7-1) and
4 beeps sound.
When an internal operating error occurs during
operation, symbol 1 (Chap XVII - Symbols) lights up
(Fig. 8-1) and 4 beeps sound. We recommend that you
IRRIGATION
FLOW
RATES
D1
60ml/min
D2
60ml/min
D3
60ml/min
D4
60ml/min
for a while (Fig. 3-4). The
MAINS
FUNCTIONS
Osteotomy,
osteoplasty
Soft tissue
removal
(Fig.5-3).