EN – Operating Manual
Due to changes in the product, the procedures stated
can change. This can take place without prior notice
or adaptation of the operating manual and is no cause
for complaint, return or price reduction. The user is
responsible for recognizing discrepancies between
the operating manual and the actual handling of the
breathalyzer and must react accordingly.
ONLY USE ORIGINAL ACE INSTRUMENTS
MOUTHPIECES
Guarantees test result accuracy
Guarantees hygiene of you and the device
Guarantees full warranty of this device
ACE INSTRUMENTS ADVISES
even if you get a result meaning „safe range", you
should never drive after consuming alcohol or do any
other activities that are severely impaired by alcohol
consumption and might endanger your health and safety
or the health and safety of others.
DISPLAY INDICATIONS
DISPLAY
MEANING
H
This letter means that
the device detected
an out of range high
alcohol concentration
(= above 2.00‰
(g/L) / 0.200% BAC /
1.000mgL).
FLo
The sampling volume
was not enough or
the blow power was
too uneven to analyze
the breath sample.
ACTION
No action required.
Please perform a
new test. Press the
start button, the
countdown will
start again. Start
breathing as soon
as "rdY" appears
on the display.
23
Snr
This means that the
sensor is not ready
for a breath sample.
This happens if high
concentration of
alcohol residue or
pollutant is found on
the sensor after the
last test.
rPt
The device could
not complete the
test successfully.
Another cause could
be the environmental
surrounding
temperature is high.
Out
The testing cannot
be performed
because the device
is turned on outside
of the operating
temperature.
Lo
The testing cannot be
performed because
the battery is low.
24
Switch off the
device and wait
at least 5 minutes
before starting it
again.
Please press the
Power button
and perform a
new test. Repeat
the test when the
environmental
surroundig
temperature is
lower.
For optimal
accuracy, repeat
the test after
the unit was
brought to a
room temperature
environment (10°C
up to 40°C).
Replace the
battery with 1 new
AA Alkaline battery.