Télécharger Imprimer la page

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10
Error
Unclosed Lid
Fan not working
Display screen shows
incomplete or
unusual characters
No display on the screen
One or several tubes are
not sealed tightly
Unit powers on with
a loud "BANG"
Unit powers on but
motor does not turn
Tube leakage
DO NOT: attempt to service the Bead Mill in a manner other than those discussed
in this manual. For any issue that is unsuccessfully corrected using this guide, please
contact your authorized dealer or call VWR 1.800.932.5000 | vwr.com
7

Troubleshooting

Cause
The lid is not latched
properly
Detection system is faulty
No Power
Faulty Fan
CPU needs to be reset
No Power
Faulty display screen
The cap is not
properly sealed or the tube
is faulty
Motor control device
damaged due to incorrect
voltage selection
Incorrect voltage selected
Speed is too high,
high temperature or
pressure inside tubes
Possible Solution
1. Check that nothing prevents the lid from
closing.
2. Press lid and ensure the handle is locked
properly.
1. Turn off the unit.
2. Contact technical assistance.
1. Check main voltage
2. Check that the voltage of the unit matches
that being delivered by the main power
supply.
3. Make sure that the unit is plugged in
properly.
4. Check and replace fuses if blown.
1. Turn off the unit.
2. Contact technical support.
1. Turn off the unit.
2. Wait 10 seconds.
3. Turn the unit back on.
1. Check main voltage.
2. Check that the voltage of the unit matches
that being delivered by the main power supply.
3. Make sure the unit is plugged in properly.
4. Check and replace fuses if blown.
1. Turn off the unit.
2. Contact technical support.
If dangerous or potentially dangerous sample
is contained in the tube, apply the proper
decontamination procedure.
Return unit to VWR for replacement
of motor control device.
Power off the unit and select appropriate
voltage. If problem persists, contact VWR for
assistance.
Reduce processing time or speed

Publicité

loading