7. Functional check

7.1. Checks

The [POWER INDICATOR] lamp indicates that high voltage is on.

7.2. Measuring efficiency with a Simco-Ion electrostatic fieldmeter

1. Take a piece of plastic foil and charge it by rubbing.
2. Measure the static charge on the plastic foil.
3. Hold the plastic in the ionized air flow at 30 cm distance for 5 seconds.
4. Measure again: The static charge should now have disappeared.

7.3. Showing efficiency, using transparent tape

1. Tear approx. 30 cm tape from the reel and hold it vertically.
2. Move your hand (or an earthed area) to the unglued side of the tape.
3. The tape is now attracted by your hand.
4. Hold the tape in the ionized air flow at 30 cm distance for approx. 5 seconds.
5. Move your hand to the tape again.
6. If the tape is no longer attracted by your hand, the static charge has disappeared.
Warning:
 Operation must not be checked by a high-voltage spark.
Long-time sparking may damage the balancing.

8. Maintenance

8.1. Summary
To ensure proper operation of the Sentry, it is essential to clean the emitter points and filter
regularly.

8.2. Cleaning the emitter points

Slide the brush several times back and forth in the outlet grating.

8.3. Cleaning the filter (optional)

1. Slide the filter out of the holder.
2. Blow-clean the filter and/or flush it with water.
3. Check that the filter is completely dry.
4. Slide the filter back into the holder.
If heavily fouled: wash out the filter, using soft soap, and flush it properly, or replace the filter.
GB
SENTRY_UM_9752066010_NL_D_GB_F_V2_0
17