Télécharger Imprimer la page

Neriox 120263 Mode D'emploi page 3

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 4
3
120263
THREE-POINT INTERNAL MICROMETER
operating instructions
1
Legend:
1. Measurement surfaces
2. Measuring head
3. Sleeve
4. Thimble
5. Lock screw
6. Ratchet
7. Ring gauge
8. Extension
9. Spanner
Attention: Ensure that the internal micrometer is checked with the aid of the supplied ring gauge prior to use.
Checking of internal micrometer
Clean the internal micrometer and the ring gauge; insert the measuring head in the ring gauge and rotate the ratchet three revolutions so that the gauge slides are firmly
fitted along the entire length. Compare scale indication with the nominal dimension of the ring gauge. The internal micrometer has to be re-adjusted unless the measurement
scale is consistent with the nominal dimension of the ring gauge.
Zero position adjustment
Clean measurement surfaces. Insert mea-
suring head in ring gauge. Rotate ratchet
three revolutions so that the gauge slides
are firmly fitted along the entire length.
a
Measurement procedure
Insert the measuring head into the hole; rotate the ratchet three revolutions so that the gauge slides are firmly fitted along the entire length. Read measured value on scale.
Reading the scale (example)
Care of instrument
We recommend using a suitable product to protect the measurement surfaces and the ring gauge against corrosion after use.
Technical data
Measurement range
mm
6...12
12...20
20...30
30...50
50...100
Repair
If the instrument is defective, please contact our service centre
Brütsch/Rüegger Werkzeuge AG, Heinrich Stutz-Strasse 20, Postfach, CH-8902 Urdorf, Tel. +41 44 736 63 63, Fax +41 44 736 63 00, www.neriox.tools, sales@brw.ch
Brütsch/Rüegger Werkzeuge AG, Heinrich Stutz-Strasse 20, Postfach, CH-8902 Urdorf, Tel. +41 44 736 63 63, Fax +41 44 736 63 00, www.neriox.tools, sales@brw.ch
3
2
8
Loosen lock screw with a hex
Allen key.
b
Measured value: 12.285 mm
Graduation
mm
0.001
0.005
0.005
0.005
0.005
Servicecenter@brw.ch
5
4
6
Rotate the sleeve until the sleeve reference
line is consistent with the "0" scale line of the
thimble. Adjust the sleeve in axial direction
so that the scale value is consistent with the
nominal value of the ring gauge.
c
Max. error of measurement
mm
0.005
0.005
0.005
0.005
0.005
7
9
Use the hex Allen key to fix the lock
screw in place at the set position.
d
Measured value: 15.715 mm
Measurable hole depth
(without extension) mm
55
80
90
97
115

Publicité