Be sure the conductivity electrode is securely attached to the meter and place it in the defined
calibration standard, wait 10-15s, and press
The calibration icon
icon
is blinking during calibration measurement. The meter reaches the end points according to
the preselected endpoint mode .
When the calibration is finished, the standard value is displayed and stored; the measurement
icon
blinks 3 times and disappears. Then, the endpoint stability and auto endpoint icon
blinks 3 times and freezes on the display. If the calibration was ended manually (by pressing
), then only endpoint stability icon
To finish the calibration and return to the sample measurement, press
then shown on the display for 3 seconds.
To reject the calibration, press
screen.
Note:
To ensure the most accurate conductivity readings, you should verify your cell constant with a
standard solution before measurement and recalibrate if necessary. Always use fresh standards.
For STCON3 the normal cell constant range is 1.50 – 2.00. (e.g. 1.71 /cm) if the cell constant is
outside the range due to an improper calibration, you may need to recover factory settings (see
5.4) , then repeat calibration.
4.2

Sample measurement

Place the conductivity electrode in the sample, wait 10-15 seconds, then press
measurement. The measurement icon
blinking during measurement and the display shows the conductivity value of the sample. When
measurement reading reaches endpoints (auto endpoint or manual endpoint); you can record the
endpoint reading result.
By pressing and holding
Stability criterion for conductivity measurement: The sensor input signal of the meter may not
change by more than 0.4% from the measured average conductivity of the probe in 6 seconds.
4.3

TDS measurement

To perform a TDS measurement, follow the same procedure as for a conductivity measurement.
Press
to switch between conductivity and TDS measurement mode.
and the measurement icon
before the meter returns automatically to the measurement
, you can switch between the auto and manual endpoint modes.
.
appear on the display. The measurement
blinks 3 times and freezes on the display.
appears on the display. The measurement icon is
EN
-
. The cell constant is
to start the
9