SYSTEM
INFORMATION
MT10 Instruction for Use - English
Date: 2012-05-01
REFLEX LEVELS
Optional:
REFLEX
FREQUENCIES
REFLEX THRESHOLD
REFLEX AUTO-STOP
REFLEX FILTER
PRINTER
BATTERY TYPE
POWER-OFF DELAY
LCD CONTRAST
EAR SEAL CHECK
REPORT CAL. DATES
SET DATE FORMAT
HOSPITAL NAME
DEPARTMENT
RELOAD DEFAULTS
SELECT LANGUAGE
Page 9/16
Select the maximum tone level to be used for the
reflex test. Set to 100 dB (with 5 dB or 10 dB steps)
or 95 dB, 90 dB or 85 dB with 5 dB steps.
Choose to perform the reflex test at 1 KHz only or
500, 1000, 2000 and 4000.
Select the change in compliance that determines
that a reflex has been detected. Adjustable in 0.01
ml steps from 0.01 to 0.5 ml.
Default 0.03 ml.
If selected, reflex measurement at each frequency
stops as soon as a reflex is found. Default YES.
Select either 2 Hz or 1.5 Hz. The lower value
smoothes the plot more.
Sanibel MPTII
Select Alkaline or NiMH (This effects the battery
state display and low battery warning).
The time before the unit turns off automatically if no
key is pressed. Select 90 or 180 seconds.
Change the display contrast 0-15. Default 7.
Select "QUICK" or "THOROUGH".
Select "PRINT CAL. DATES" or "HIDE
CAL.DATES".
Select "DD/MM/YY" or "MM/DD/YY"
Allows the hospital name to be entered (this will
appear at the top of the print out).
Allows the department name to be entered (this will
appear at the top of the print out).
The options above are reset to their default values.
Select "ENGLISH", "GERMAN" or "FRENCH" for
operating language.
Shows: Battery voltage
Software version
Date calibrated
Next calibration date
Instrument serial number
Current date and time.