Technical Specifications - Beurer medical BC 80 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26

10. Technical specifications

Model no.
BC 80
Measurement method Oscillometric, non-invasive blood pres-
sure measurement on the wrist
Measurement range
Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 97 mm x W 68 mm x H 22 mm
Weight
Approximately 108 g (without batteries)
Cuff size
135 to 230 mm
Permissible operating
+5 °C to +40 °C, 15-93 % relative hu-
conditions
midity (non-condensing)
Permissible storage
-25 °C to +70 °C, 10-93 % relative hu-
conditions
midity, 700-1050 hPa ambient pressure
Power supply
2 x 1.5V
Battery life
For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
AAA batteries
25
Accessories
Instructions for use, 2 x 1.5V AAA bat-
teries, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
• This device is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to electro-
magnetic compatibility (EMC). Please note that portable and
mobile HF communication systems may interfere with this
unit. More details can be requested from the stated Customer
Services address or found at the end of the instructions for
use.
• This device corresponds to the EU Medical Devices Direc-
tive 93/42/EEC, the German Medical Devices Act (Mediz-
inproduktgesetz) and the standards EN 1060-1 (non-inva-
sive sphygmomanometers, Part 1: General requirements),
EN  1060-3 (non-invasive sphygmomanometers, Part3:
Supplementary requirements for electro-mechanical blood
pressure measuring systems) and IEC 80601-2-30 (Medical
electrical equipment – Part 2-30: Particular requirements for
the basic safety and essential performance of automated
non-invasive sphygmomanometers).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful life.
If using the device for commercial medical purposes, it must
be regularly tested for accuracy by appropriate means. Pre-
cise instructions for checking accuracy may be requested
from the service address.

Publicité

Table des Matières
loading

Table des Matières