Result Number
POSITIVE CONTROL LINE
02
CONSULT MANUAL
03
TEST DEVICE
BACKGROUND INVALID
05
CONSULT MANUAL
06
07
08
09
11
NEGATIVE CONTROL
12
LINE INVALID
CONSULT MANUAL
13
14
Message
Possible Cause(s)
The positive assay control
INVALID
line is outside the normal
range.
TEST WAS
The test device was removed
REMOVED
while the run was in
REINSERT
progress.
The assay background area
is outside of the normal
ERROR 05
range.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 06
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 07
ANALYZER
A high ambient light
FAILURE
condition has been detected
CALL BD
or a hardware anomaly has
ERROR 08
occurred.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 09
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 11
Cross-reacting material
present.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 13
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 14
Corrective Action(s)
On first occurrence, repeat the assay. On subsequent
occurrences, contact BD.
That run is invalidated. The device may be re-inserted to
begin a new run.
Remove the device and cycle the instrument power.
If the instrument displays an error during power-on,
contact BD.
If no error message appears, obtain an additional
sample and repeat the assay. If the issue recurs,
contact BD.
Contact BD.
Contact BD.
On first occurrence, move the unit to a lower ambient light
condition.
In case of persistent recurrence, contact BD.
Contact BD.
Contact BD.
On first occurrence, repeat the assay. On subsequent
occurrences, contact BD.
Contact BD.
Contact BD.
31