Autoclaving

The Transferpette
(121 °C) and 20 min. exposure time ( t
1.
Eject the pipette tips.
2.
Autoclave the complete pipette without
any further disassembling.
3.
Allow the Transferpette
completely cool and dry.
Note:
The effectiveness of the autoclaving must
be verified by the user.
Maximum reliability is obtained with vacuum
sterilization. We recommend the use of
sterilization bags.
Attention:
During autoclaving, the volume
setting lock must be in the upper
switch position (UNLOCK)!
If the pipette is autoclaved frequently, the
pistons should be oiled with the supplied
silicone oil in order to preserve smooth
movement.
If necessary after autoclaving, tighten the
connection between the hand grip and the
manifold.