Accuracy Table

Final test values related to the nominal capacity (maxi-
mum volume) indicated on the instrument, obtained
when instrument and distilled water are equilibrated at
ambient temperature (20 °C/68 °F) and with smooth
operation. According to DIN EN ISO 8655.
Transferpette
Volume range
µl
0.5 - 10
5 - 50
10 - 100
20 - 200
30 - 300
* A = Accuracy
* CV = Coefficient of Variation
32
®
S -8/-12
Volume step
µl
10
5
1
50
25
5
100
50
10
200
100
20
300
150
30
A*
CV*
Increment
≤ ± %
≤ %
µl
1.6
1.0
0.1
2
2
8
6
0.8
0.4
0.1
1.4
0.8
6
3
0.8
0.3
0.2
1.4
0.6
4
2
0.8
0.3
0.2
1.4
0.6
4
1.5
0.6
0.3
0.2
1.2
0.6
3
1.5
20 °C
!
Ex
Type of tips
µl
20
200
200 / 300
200 / 300
300