C2AI FCX-All-V5 Série Manuel D'utilisation page 81

Table des Matières

Publicité

G
OUT
B:
OUTPUT ADJ
< INC > < CHANGE >
CHNG
INC
B-1: OUTPUT SET
0mA
< LRV/URV/ENT/CL >
CL
1
LRV
URV ENT
B-1: OUTPUT SET
B-1: OUTPUT SET
4.000mA
CHANGE OK? < ENT/CL >
CHANGE OK? < ENT/CL >
2
7
CL
ENT
(4 , 20mA)
B-2:
OUTPUT MODE
XX.XXXmA
CURRENT CONST.
< CALB > < CHNG > < CL >
3
CL
CALB
CHNG
B-3: OUTPUT CALIB
XX.XXXmA CONST.
0mA (EXT.mA)
< ENT > < CL >
4
ENT
CL
B-3:
OUTPUT CALIB
XX.XXXmA CONST.
YY.YYYmA (EXT.mA)
CHANGE OK? < ENT/CL >
5
ENT
CL
B-4:
OUTPUT
XX.XXXmA CONST.
YY.YYYmA (EXT.mA)
CALIBRATING
6
Calibration
impossible
B-4:
OUTPUT
XX.XXXmA CONST.
YY.YYYmA (EXT.mA)
NOT CAL1B. < CL >
10
CL
C:
DATA
80
(Input of XX.XXXmA at
1
B-1: OUTPUT SET
20.000mA
XX.XXXmA
CHANGE OK? < ENT/CL >
ENT
CL
ENT
CL
8
(OTHERS)
< OTHERS >
B-2:
OUTPUT MODE
XX.XXXmA
CURRENT CONST.
< CHANGE > < CL >
9
CL
CHNG
appears when the
10
adjustable range is
exceeded.
(étalonnage du signal de sortie) permet l'éta-
lonnage du convertisseur numérique/analo-
le signal de sortie.
)
raccordés dans la boucle 4-20 mA.
en appuyant sur la touche <OUT
appuyant sur la touche <INC
l'étalonnage de la valeur initiale 4 mA et sur la
touche <URV
valeur de consigne correspondante apparaît
CONST." (courant constant). Le courant cor-
appuyé sur la touche <ENT
précision est en dehors des tolérances du
ner, on peut alors procéder à une correction.
par la touche <ENT
la touche <CL
Par sécurité "CHNG OK ?
doit être validé par la touche <ENT
Le transmetteur calculera alors automatique-
ment les écarts entre la valeur théorique (4 ou
20 mA) et la valeur mesurée et apportera les
corrections nécessaires au niveau du conver-
tisseur D/A. Cette procédure est applicable
aussi bien à LRV (4mA) qu'à URV (20 mA).
LRV
-
TN5FCXA2V5h-F