Actuating the harvester handle prior to placement of the 3M™ Tegaderm™ Film may result in an inability to
secure the micrografts for application.
Wipe the vacuum head and the control unit nest following institutional procedures used for cleaning
and disinfection of other hard surface durable electronic medical equipment between each patient use.
Inadequate cleaning may result in patient contamination.
Use this system only with the supplied components.
Discard all disposable items (harvester and 3M™ Tegaderm™ Film) in accordance with local medical waste
disposal regulations.
Contact KCI for disposal of the control unit at the end of its expected service life. This product is designated
for separate collection at an appropriate collection point. Do not dispose of this product in normal waste
stream.
Handle the vacuum head with care. Do not drop the vacuum head, as doing so could damage the glass
warming window. Do not suspend or hold the vacuum head by the tubing
Reversible maceration at the recipient site is a potential risk. Please use best clinical judgement for
preventing maceration. Fenestration of the 3M™ Tegaderm™ Film with an 18g needle prior to application
onto the recipient site may be appropriate.

CONTROL UNIT USER INTERFACE PANEL

Status Indicator:

steady amber - unit plugged in
steady green - unit turned on
flashing green - unit running
flashing amber - fault condition (see below)
Vacuum Fault: Indicates that sufficient vacuum to perform the procedure has not been achieved. In this
case ensure that the harvester is securely strapped onto the patient's donor site, and the vacuum head is
securely snapped onto the harvester.
Other Fault / Service Unit: Do not use the system. Contact your KCI representative.
Refer to page 10 for control unit interface panel messages and troubleshooting.
NOTE:
Power
Start / Pause
Timer Down
Timer Up
Timer Display
12
Status Indicator
Vacuum Fault
Other Fault / Service Unit