Télécharger Imprimer la page

Error Codes - IKA RCT basic safety control Mode D'emploi

Masquer les pouces Voir aussi pour RCT basic safety control:

Publicité

Error codes

Any malfunctions during operation will be identified by an error
message on the display (E and F).
Proceed as follows in such cases:
Error code Cause
E3
Temperature inside device too high
E4
Motor blockage
E6
Break in safety circuit
E24
Surface temperature (temperature of control sensor):
of the base plate is higher then the setted safe tempera-
ture limit
E44
Surface temperature (temperature of safety sensor):
of the base plate is higher then the setted safe tempera-
ture limit
E25
Heating and switching element monitoring
E26
Difference between temperature of safety sensor and
temperature of control senor
control temperature > (safety temperature + 40 K)
E46
Difference between temperature of safety sensor and
temperature of control senor
safety temperature > (control temperature + 40 K)
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short desciption of the fault.
R_HCT_042011
F Switch off device using the main switch (A)
F Carry out corrective measures
F Restart device
Effect
Solution
Heating off
- Switch off device and allow to cool down
Heating off
- Switch off device
Motor off
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in
connector for the motor
Heating off
- Plug in contact plug (L)
- Plu in PT 1000 contact thermometer / temperature sensor
- Replace faulty connecting cable, plug, or contact thermometer
Heating off
- Switch off device until the surface temperature of the base
plate is lower than the selected safe temperature limit
- Set ahigher safe temperature limit
Heating off
- Switch off device until the surface temperature of the base
plate is lower than the selected safe temperature limit
- Set ahigher safe temperature limit
Heating off
- Switch off device
- Set the safe temperature limit > 55 °C
see also „Fuctional check of inactivating the safety circuit"
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in
connector for the heating element
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in
connector for the temperature sensor
Heating off
- Switch off device
- Warning! Only to be carried out by authorised service personnel:
Carry out an internal test on the device to check the plug-in
connector for the temperature sensor
21

Publicité

loading

Ce manuel est également adapté pour:

Hct basic safety control