Télécharger Imprimer la page

Agilent Technologies TV 401 Notice D'emploi page 239

Publicité

N.
402
404
500
503
504
1. Automatic means that when the controller stops, the vent valve is opened
NOTE
with a delay defined by window n. 126; when the controller starts, the vent
valve is immediately closed.
2. On command means that the vent valve is opened or closed by means of
window n. 122.
3. These commands remove the pump motor supply for a short time. If they are
used continuously, they can cause the pump stopping.
Figure 25
TV 401/301 MS User Manual / 87-901-013-01 (B)
Read/
Data
Write
Type
R
A
CRC Param. (PA)
R
A
CRC Parameter structure
Reserved to Agilent service
R/W
N
RS 485 address
R/W
L
Serial type select
7
6
TOO HIGH LOAD
SHORT CIRCUIT
OVERVOLTAGE
AUX FAIL
Window N. 206 Bit Description
Description
5
4
3
2
1
0
NO CONNECTION
PUMP OVERTEMP.
CONTROLL. OVERTEMP.
POWER FAIL
Technical Information
Window-Meanings
Admitted Values
PA8XXXX
(where "XXXX" are
variable)
"XXXX"
0 to 31
(default = 1)
0 = RS 232
1 = RS 485
(default = 0)
239/266
17

Publicité

loading

Ce manuel est également adapté pour:

301 msTv 401/301/21 nsf