CLEANING AND MAINTENANCE

AIR FILTER
Your Ombra* Portable Compressor is equipped with a special air filter that will slowly change colour with use. The
original filter is bright white, but it will slowly turn grey with use. Your air filter should be changed every 30 days, or as
visual inspection indicates. The compressor should never be used without an air filter installed.
To change the filter (Figure 8)
1. Pull back the
AIR FILTER COVER
2. Dispose of the old filter.
3. Insert a new filter, then replace
CAUTION: Do not wash or reuse the filter. Use only Ombra* Replacement Filters (Part Number: 10554493050).
Do not attempt to substitute any other material to act as a filter as this will result in damage to the compressor.
FIGURE 8
NEBULIZER
Refer to the Owner's Manual included with your reusable AEROECLIPSE* XL BAN for instructions on how to properly
clean your nebulizer.
EZ TWIST TUBING
Occasionally you may notice condensation on the inside of your oxygen tubing. This is normal. To remove the
condensation, simply connect one end of the EZ Twist
compressor to run without the nebulizer attached. This will force air through the tubing, drying the inside. To clean the
exterior, wipe with a damp cloth. Ensure the tubing is dry prior to storage.
COMPRESSOR
Prior to cleaning the compressor, ensure the unit is unplugged from the electrical outlet. Use a damp cloth to wipe the
exterior of the compressor housing. Never use liquid or spray cleaners directly on the compressor. Ensure liquids or
moisture do not get onto the battery
could cause damage.
CAUTION: Liquids can cause damage to the electrical components inside your compressor and lead to malfunction.
In the event any liquid enters the unit, please contact your Ombra* Authorized Distributor. Do not use the compressor
until it has been inspected, even if it appears to be functioning properly.

RECHARGEABLE BATTERY

If the Ombra* Rechargeable Battery requires cleaning, use only a damp cloth to wipe the exterior surfaces. Ensure no
moisture or liquids go near the battery
NOTE: Do not leave the Ombra* Rechargeable Battery plugged in to an electrical source for more than 7 days.
Leaving the battery connected to a power source for long periods of time may decrease the life of the battery. The battery
should be charged using the AC Adapter for 6 hours at least every 4 months to optimize and preserve performance.
and remove the old filter from the compressor.
FILTER
COVER.
TUBING
CONTACT POINTS
on either the Compressor or the Rechargeable Battery as this
CONTACT
POINTS.