I-Tech MOI-IONOTENS Mode D'emploi page 25

Table des Matières

Publicité

Electrostatic
discharge
(ESD)
EN 61000-4-2
Mains
power
electromagnetic field
EN 61000-4-8
Guidance and manufacturer's declaration – electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that is used in suche environment
Immunity test
Test level EN
Conducted RF
3 Vrms 150kHz
EN 61000-4-6
to 80MHz
RF Radiata
3 Vrms 80MHz
EN 61000-4-3
to 2,5GHz
Field strangths from fixed RF transmitters, are determined by an electromagnetic site survey, should be less than the
complicance level in each frequency rage.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended separation distances between portable and mobile communications equipment
The device is intended for the use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the device can help prevent electromagnetic interferences by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the device as recommended
below, according to the maximum output power of the communication equipment.
Rated
maximum
power
of
the
transmitter (W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
Note:
(1) At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
IACER S.r.l.
 6kV contact
 6kV contact
 8kV air
 8kV air
3 A/m
3 A/m
Compliance level
60601-1-2
3 Vrms 150kHz to
80MHz
3 Vrms 80MHz to
2,5GHz
and the device
Separation distance according to the frequency of the transmitter (m)
150kHz to 80MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
25 di 28
Floors sholud be wood, concrete or
ceramic tile. If floor are covered with
syntethic material, the relative
humidity should be at least 30%
Mains power quality should be at that
of a typical commercial or hospital
environment
Electromagnetic environment –
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
d = 1,2 P 150kHz to 80MHz
d = 1,2 P 80 MHz to 800 MHz
d = 2,3 P 800 MHz to 2,5 GHz
where (P) is the maximum output power rating
of the transmitter in Watts (W) according to the
transmitter manufacturer and (d) is the
recommended separation distance in metres
(m).
80MHz to 800MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
MNPG170 Rev. 05 11/11/15
800MHz to 2GHz
d = 2,3 P
0,23
0,73
2,3
7,3
23

Publicité

Table des Matières
loading

Table des Matières