Installation
Before connecting the aspirator to the piping of the centralized
system, ascertain that aspiration piping is clean as heavy debris
could damage the appliances.
Connect the PVC light grey aspiration tube (2b) (supplied with
the machine) to the 50 mm Ø tube-holder (2) ("aspirated fl uid
inlet"). The other end of the same tube must be connected to the
aspiration piping (3) coming from the surgeries (page 18).
The black heat resistant exhaust air pipe (4b), fi tted with a
metal spiral, must be connected to the 50 mm Ø tube-holder
(4) ("exhaust air outlet"). Connect the other end of the pipe to
the antibacterial fi lter (5), passing preferably through a silencer
(5a) supplied with the aspirator. The hot air coming from the
antibacterial fi lter must be conveyed outside.
Connect the 18 mm Ø tube-holder (6) to the liquid drain pipe.
In the version fi tted with Hydrocyclone the aspirator is draining
liquids by gravity, therefore these can never be drained towards
the top.
The 10 mm Ø tube-holder (8) ("emergency drain")
must be connected to the liquid drain pipe.
The piping connecting the machine to the aspiration
and draining system should be fl exible to damper the
small vibrations produced by the aspirator.
The aspiration piping should be run in the fl oor and at
a point near the aspirator it should rise about 30 cm to
reach the tube-holder (2) (Draw. A and B, page 61).
If Turbo HP or Turbo-Jet modular is installed on a
lower fl oor than the dental units, the aspiration
piping must not enter the centrifugal separator
perpendicularly. Rather, place a few metres of piping
horizontally on the same level of the machine and then
install it at the centrifugal separator inlet which is on a higher
level (use a fl exible pipe) (Draw. B, page 61). The aspirated fl uid
is reaspirated by the machine.
After the installation, connect the power cable to the mains accord-
ing to the regulation EN 61008-1. Finally connect the low voltage
users line: the clamps 1-2-3 of the circuits AC
48B/AC58B and AC58 (Draw. F-G-H, pages 67-
68-69) to the users line (12) (page 19 and Draw. C,
page 62) which is feeding all the dental units of the
same system with connections in parallel. Ascertain
that the contact on the dental unit is clean (free
contact without voltage).
19
4b
18
2b
14
4
5a
6
2
8
2b
3
5