8 Maintenance

8.3 Sending in parts

8.4 Replacing the disposable filter

Applies to
Requirement
28
EN
OXYWAY
Risk of infection due to contaminated parts during
maintenance work!
The device, accessories and other parts may be contaminated and
infect the technicians with bacteria and viruses. Sent in parts
which are visibly contaminated will be disposed of by
WEINMANN Emergency or personnel authorized by
WEINMANN Emergency at the expense of the sender.
 Clean and disinfect the parts required for maintenance
measures.
 Do not send in parts which are potentially contaminated.
1. Disassemble parts.
2. Clean and disinfect parts
page
21).
3. Send in the parts to WEINMANN Emergency or to a technician
specifically authorized by WEINMANN Emergency.
Pressure reducer without PIN index connection
The pressure reducer is removed
pressure reducer", page
Wash your hands thoroughly in advance.
The Allen key (5 mm) must be clean, dry and free from any oil
or grease.
(see "5 Hygienic reprocessing",
(see "4.2 Removing the
16).