3. Remove the low-pressure hose from the pressure outlet.
or
If the device has an inhalation outlet: Remove the accessories.
4. Remove the hand-screwed connection from the valve of the
oxygen cylinder by hand.
Result
The pressure reducer is removed.

4.3 Inserting the pressure reducer

Risk of injury from insufficient oxygen supply!
In cases with an oxygen requirement of more than 90 l/min, there
is a risk of an insufficient oxygen supply, which puts the patient at
risk if the pressure outlet and inhalation outlet are running at the
same time.
 Do not use the pressure outlet and inhalation outlet at the
same time.